Set the electrode layout, wiring and mounting dimensions. Review the section drawing or export it at full scale for your CAD system.

Measurement

Sigma · TX/RX conductance. Separate transmitter and receiver.

Section dimensions

Each axial channel repeats the electrode layout at a new position along the pipe.

Flange and body dimensions
View updated drawing ↑
Process and acquisition
Flange pressure reference

No tabulated pressure reference for the selected material, interface and temperature range.

Assembly pressure and temperature limits require review of seals, body and feedthroughs. The ASME screening table covers 0–200 °C.

Structure velocity uncertainty

Track the same interface at two sensing positions. Use its expected speed and the distance between their centres to plan the measurement.

Expected transit time
70 ms
Samples across transit
700
Sampling contribution to delay (k = 1)
0.04082 ms
Standard velocity uncertainty (k = 1)
0.0005832 m/s · 0.05832%
Expanded velocity uncertainty (k = 2)
±0.001166 m/s · 0.1166%

First-order design estimate for v = L/Δt. Blank optional terms contribute zero. k = 2 is a coverage factor.

Calculation and assumptions

u(v)/v = √[(u(L)/L)² + (u(Δt)/Δt)²]. Each arrival is picked to the nearest sample. Independent, uniformly distributed rounding errors give a standard delay uncertainty of 1/(√6 × sampling rate). Spacing and additional delay uncertainties are independent standard values (k = 1). Shared-clock rounding can be correlated. Detection bias, flow evolution and clock errors require experimental characterisation. Uncertainty propagation · NIST